TM 55-4920-401-13&P
Table 3.2. Troubleshooting Table
Malfunction
Probable Cause
Corrective Action
1 . I N S U L A T I O N C H E C K m e t er
a . M e c h a n i c a l z e r o
a. Adjust screw.
d o e s n o t r e a d i n f i n i t e r e -
s c r e w o u t o f a d j u s t -
sistance when FUNCTION SELECT
ment
switch is off
2 . R E S I S T A N C E C H E C K m e t er
a . M e c h a n i c a l z e r o
a. Adjust screw.
d o e s n o t r e a d 0 w h e n F U N C -
s c r e w o u t o f a d j u s t -
TION SELECT switch is off
ment
3 .
E n t i r e t e s t e r i n o p e r a -
a . N o i n p u t v o l t a g e
a. Check power source.
tive with power source connected
b. Defective power cable
b. Replace cable.
a n d M A S T E R P O W E R s w i t c h
c . D e f e c t i v e p o w e r
c. Replace cable.
in ON position.
interconnect cable
4 . P O W E R O N l a m p d o e s n o t
a. Defective POWER ON
a. Replace lamp (para.
light when MASTER POWER switch
lamp
3-12).
is on.
5 . P o w e r i s a p p l i e d t o t e s t er
a. DISPLAY control in
a. Turn control clock-
but displays do not light.
f u l l c o u n t e r c l o c k w i s e
wise.
position
6 . A l l d i s p l a y d i g i t s d o n o t
a . D e f e c t i v e d i s p l a y
a. Replace display
read 8 when PRESS TO TEST button
on display board
(para. 3-13).
is pressed.
7 . S T D D A Y l a m p d o e s n o t
a . D e f e c t i v e S T D
a. Replace lamp (para.
l i g h t w h e n S T D D A Y s w i t c h is
DAY lamp
3-12).
on
8. PROBE POWER lamp does not
a . D e f e c t i v e P R O B E
a. Replace lamp (para.
l i g h t w h e n h e a t e r p r o b e s a r e
POWER lamp
3-12).
c o n n e c t e d , F U N C T I O N S E L E CT
b. FUSE 20A blown
b. Replace fuse (para. 3-14).
s w i t c h i s i n H E A T E R P R O B E
position, T/C INPUT switch is in
H E A T E R C A B L E p o s i t i o n , a nd
P R O B E C O N T R O L i s t u r n e d
clockwise.
9. No heater probes heat.
a. Defective heater cable.
a. Replace cable.
10. One of several heater probes
a. Defective heater probe
a. Replace probe.
does not heat.
b. Defective junction box
b. Replace box.
11. INSULATION CHECK meter reads
a. Defective insulation
a. Replace adapter.
i n f i n i t y
w h e n
i n s u l a t i o n
check adapter
check adapter is jumpered.
b. Defective instrument
b. Replace cable.
cable
3-5