Table 3-2. Troubleshooting - continued.
MALFUNCTION
TEST OR INSPECTION
CORRECTIVE ACTION
13.
FLASH
Step 1.
Step 2.
tripped.
POINT TESTER HEATING ELEMENT FAILS TO HEAT AIR BATH.
Check the rheostat control knob is mounted to shaft correctly.
Tighten rheostat control knob.
Check and reset circuit breaker for wall receptacle which tester is connected is not
If flash point tester heating element still fails to heat, notify unit maintenance.
POINT TESTER STIRRER FAILS TO OPERATE
14.
FLASH
Check that power cord is plugged in and line switch is on.
a.
Plug in power cord and turn on line switch.
b.
If flash point tester stirrer still fails to operate, notify unit maintenance.
15.
DISTILLATION UNIT FAILS TO HEAT FLASK CONTAINING TEST SAMPLE.
Step 1.
Step 2.
Step 3.
Check that power cord from shield assembly is plugged in to power receptacle.
Plug in power cord.
Check that power cord from shield assembly is connected to condenser assembly.
Check and reset circuit breaker CB12 for wall receptacle in which distillation unit
is plugged into is not tripped.
If distillation unit still fails to heat flask, notify unit maintenance.
16.
DISTILLATION UNIT CONDENSER ASSEMBLY FAILS TO MAINTAIN PROPER
TEMPERATURE.
Step 1.
Verify power cord from shield assembly is plugged into power receptacle.
Plug in power cord.